Ultrasonic Impedance Analysis Machine For Measuring Frequency Of Ultrasonic Transducer Testing The Parameters

17 - Ultrasonic Impedance Analysis Machine For Measuring Frequency Of Ultrasonic Transducer Testing The ParametersImpedance tester application:

 

Transducers: The admittance circle diagram should not show parasitic circles, the resonant frequency should be as close as possible to the design frequency, the dynamic impedance should be low, the quality factor Qm should be close to the design requirements, and the capacitance should match the circuit.

 

Piezoelectric ceramic sheet: The device can be directly judged based on the admittance circle diagram and logarithmic coordinates. If there are layers or cracks inside the ceramic sheet, the logarithmic curve will have multiple peaks, and multiple parasitic small circles will appear on the admittance circle diagram.

 

The design, processing, and assembly of the horn are reasonable or have defects, which can be clearly seen on the guide circle diagram.

Product name Sound intensity measuring instrument Numbering
Model PLS-SYJ100 Date of inspection
Testing base Enterprise production specification Detection parameter The test sound intensity is 0.8W/cm², which meets the factory requirements
Test equipment parameters 빈도: KHz로, 힘: KW.
Test items Sound intensity meter parameter measurement Conclusion Compliance with the factory-leaving standard of enterprises