Impedance tester application:
Transducers: The admittance circle diagram should not show parasitic circles, the resonant frequency should be as close as possible to the design frequency, the dynamic impedance should be low, the quality factor Qm should be close to the design requirements, and the capacitance should match the circuit.
Piezoelectric ceramic sheet: The device can be directly judged based on the admittance circle diagram and logarithmic coordinates. If there are layers or cracks inside the ceramic sheet, the logarithmic curve will have multiple peaks, and multiple parasitic small circles will appear on the admittance circle diagram.
The design, processing, and assembly of the horn are reasonable or have defects, which can be clearly seen on the guide circle diagram.
产品名称 | 声强测量仪 | 编号 | |
型 | PLS-SYJ100 系列 | 检查日期 | |
检测基地 | 企业生产规范 | 检测参数 | 测试声强为 0.8W/cm², 符合工厂要求 |
测试设备参数 | 频率: 千 赫, 权力: 千瓦. | ||
测试项目 | 声强计参数测量 | 结论 | 符合企业出厂标准 |