Ultrasonic power meters are classified according to measurement principles: Currently, the measurement methods for ultrasonic power mainly include radiation force method, electrical method, light diffraction method, and calorimetry method. In ultrasonic power measurement, the radiation force method can directly measure the total ultrasonic radiation power value without considering the limitations of ultrasonic entry and far-field. Therefore, it is the most widely used and is also the preferred sound power measurement method recommended by IEC.
The basic principle of the radiation force method is to calculate the sound power emitted by the ultrasonic transducer by measuring the radiation pressure acting on the obstacle (target) in the measured sound field.
Electronic balance measurement of sound radiation force method:
W=cF/(1+r2cos2) φ)
W – Total sound power, unit w
C – The propagation speed of ultrasound in liquid, in m/s
F – The force acting on the target along the ultrasonic axis direction, in N
Φ— The angle between the normal of the target surface and the incident sound beam, in degrees
產品名稱 | 聲強測量儀 | 編號 | |
型 | PLS-SYJ100型 | 檢查日期 | |
檢測基地 | 企業生產規範 | 檢測參數 | 測試聲強為0.8W/cm², 符合工廠要求 |
測試設備參數 | 頻率: 千赫, 權力: 千瓦. | ||
檢測項目 | 聲強計參數測量 | 結論 | 符合企業出廠標準 |